Positive pressure tester for leakage of infusion set
Product introduction:
Applicable standards: the medical device sealing tester is a
special equipment designed according to the relevant provisions in
the national standards for medical devices, which is specially used
to test the sealing performance of medical devices.
Positive pressure output range: 20KPa ~ 300kPa (adjustable), LED
digital display, the error is not more than ± 5%.
Keep the positive pressure of the measuring system: do not drop
1kPa within 1 minute.
Test duration: 1s ~ 60s (adjustable), LED digital display, error no
more than ± 1s.
Pressure output range: 20 kPa ~ 300 kPa higher than local
atmospheric pressure,
Display mode: LED digital display; pressure output value can be set
arbitrarily, and the calibration error is not more than ± 2.5% of
the reading.
Test duration: 1s ~ 99.9 minutes, LED display, error is not more
than ± 1s.
It is suitable for infusion set, blood transfusion set, infusion
needle, anesthesia bag filter, pipeline, catheter, quick connector,
etc.
External air compressor:
It can work continuously for at least 16 hours a day;
External compressed air (instead of air compressor) interface is
available for online positive pressure detection of various medical
devices;
Zsj8368-b infusion set injection parts test device
It is used with sz8368-b infusion set leakage positive pressure
tester
Speed of puncture and cotton cloth movement: 500mm / min
Time controller: 1s ~ 60s (adjustable), LED digital display, the
error is not more than ± 0.1s.
When the pressure is 50 kPa higher than the atmospheric pressure,
the injection part is punctured with an injection needle with an
outer diameter of 0.8 mm. After 15 s of insertion, there is no
leakage within 1 min after pulling out.
![]() |